Will three radial hinges and a low-profile brass threshold for doors be acceptable? The brass threshold is acceptable. Regarding the hinges, the design intent for the continuous hinge is a) greater durability, and b) to help insure a greater RF seal. All doors during Shop Drawing reviews will be reviewed against this standard."
.... Read More
|
||
| Office Location | Agency Name | Solicitation Base Posting Type |
| VETERANS AFFAIRS, DEPARTMENT OF | Combined Synopsis/Solicitation | |
| SetAside Type | ||
| Solicitation ID | Solicitation Title | Solicitation Office |
| 36C24626B0001 | Z2DA--Replace MRI 1.5T | |
| Synopsis | ||
|
It calls out fiber for the injector to the actual MRI machine. Need to know if it is Sigle Mode or Multi Mode fiber. How many strands of fiber they are needing there and the type of connectors they need on the end of the fiber. Injector has a fiber optic cable that is supplied with the purchase of the injector, so the contractor is not responsible for providing that cable. There is no additional fiber needed for the MRI machine. Need to know how many data drops per location it is not called out anywhere also need to know .... Read More
|
||
| Office Location | Agency Name | Solicitation Base Posting Type |
| VETERANS AFFAIRS, DEPARTMENT OF | Combined Synopsis/Solicitation | |
| SetAside Type | ||
| Solicitation ID | Solicitation Title | Solicitation Office |
| 36C24626B0001 | Z2DA--Replace MRI 1.5T | |
| Synopsis | ||
|
7. Do we know that Siemens has confirmed that there will be no active interference between the two MRI systems? We have the ability to quote magnetic compensation but need to know if it's necessary. The contractor will not be responsible for any magnetic compensation 25. PACS: Can we the government please provide existing control panels and headend software id info, licensing info and photos? The PACS is primarily controlled by a Phillips Filter Box in the Mechanical room behind the MRI machine. The Filter Box works in conjunction with a control panel located in the control room at the .... Read More
|
||